CO-oximeter
A CO-oximeter is a device that measures the oxygen carrying state of hemoglobin in a blood specimen including oxygen-carrying hemoglobin (O2Hb), non-oxygen-carrying but normal hemoglobin (HHb) (formerly, but incorrectly referred to as 'reduced' hemoglobin) as well as the dyshemoglobins such as carboxyhemoglobin(COHb) and methemoglobin(MetHb). The use of 'CO' rather than 'Co' or 'co' is more appropriate since this designation represents a device that measures carbon monoxide (CO) bound to hemoglobin, as distinguished from simple oximetry which measures hemoglobin bound to molecular oxygen- O2Hb (or capable of binding to molecular oxygen-HHb. The simpler oximeters may report oxygen saturation alone which is the ratio of oxyhemoglobin to total 'bindable' hemoglobin (i.e. oxyhemoglobin + deoxyhemoglobin-HHb).
CO-oximetry is useful in defining the causes for hypoxemia, or hypoxia, (oxygen deficiency at the tissue level). The reporting of multiple hemoglobin moieties or fractions is done with a device that measures absorption of light passing through blood at as few as two or three wavelengths of light to as many as several dozens of wavelengths to distinguish oxyhemoglobin, and deoxyhemoglobin (formerly called 'reduces' hemoglobin, and thus determine the oxyhemoglobin saturation (the percentage of oxygenated hemoglobin compared to the total amount of available hemoglobin (Hb)). Measurement of greater numbers of wavelengths enables the instrument to distinguish between these and carboxyhemoglobin,-COHb, methemoglobin -metHb, other hemoglobin moieties and 'background' light-absorbing species.
When a patient presents with carbon monoxide poisoning (CO) or other non-respiratory hypoxic symptoms, most current CO-oximeter will detect the relative levels of each hemoglobin fraction (oxyhemoglobin and dyshemoglobins) and likely the oxyhemoglobin saturation. For any system making these measurements it is critical that the device clearly distinguish between Oxygen Saturation' and Fractional Oxyhemoglobin" . The issue here is the careless use of saturation vs. fractional oxyhemoglobin, which both measure the same entity -oxyhemoglobin- but the oxygen saturation uses as its base only the hemoglobin available for binding, while the fractional oxyhemoglobin uses the total hemoglobin in the sample as its base. In normal subjects the values are nearly identical-thus leading to terminologic and possibly clinical confusion. A simple oximeter measuring only oxygen derivatives, may report a normal saturation or even a hyperoxic state if oxygen gas has been administered when in fact there is serious compromise of oxygen carrying ability of the hemoglobin present..
Measurement
Traditionally, this measurement is made from arterial blood processed in specific device designed to be able to measure proportions of multiple components of several hemoglobin moieties using multi-wavelength spectrophotometry and complex, but straightforward internal computations. While these units still are in wide use, blood gas analyzers with integral CO-oximetry modules have also been developed and successfully marketed by several manufacturers.[1][2] More recently, some 'pulse' or more precisely 'peripheral' oximeters have made it possible to estimate carboxyhemoglobin with non-invasive technology similar to a simple (peripheral) pulse oximeter.[3] In contrast, the use of a standard or simple pulse oximeter is not effective in the diagnosis of CO poisoning as patients suffering from carbon monoxide poisoning may have a normal oxygen saturation reading on a pulse oximeter.[4]
References
CLSI, C46-A2- Blood Gas and pH Analysis and Related Measurements; Approved Guideline—Second Edition, Wayne, PA, 2010
Zijlstra WJ, Maas AHJ, Moran RF. Definition, significance and measurement of quantities pertaining to the oxygen carrying properties of human blood.. Scand J Clin Lab Invest , 56(Suppl), 224, 27-45, 1996
Brunelle JA, Degtiarov AM, Moran RF, Race LA, Simultaneous measurement of total hemoglobin and its derivatives in blood using CO-oximeters: Analytical principles; Their application in selecting analytical wavelengths and reference methods; A comparison of the results of the choices made. Scand J Clin Lab Invest, 56: (Suppl) 224, 47-69, 1996.
Brunelle JA, Moran RF, Data processing in CO-oximeters that Use overdetermined systems (Reply). Clin Chem, 43:1, 189-191, 1997
Degen BR, Moran RF, Comparison and assessment of blood gas related quantities including base excess, the gas exchange indices and temperature corrected pH/ PO2/PCO2 , as defined in approved NCCLS standard C12-A, using a computer simulation of input variables., Scand J Clin Lab Invest , 56:(Suppl) 224, 89-106 1996.
Moran R, Hemoglobin F and measurement of oxygen saturation and fractional oxyhemoglobin. Clin Lab Sci, 7:3, 162-164, 1994.
Brunnelle JA, Degtiarov AM, Moran RF, Race LA, CO-oximetric measurement of oxyhemoglobin, deoxyhemoglobin, and dyshemoglobins in blood: Effects of analytical wavelength and reference method selection. Lab Hematol. 1:2, 161 - 164, 1995.(
Moran RF, Implications of Fetal Hemoglobin : Measurement of oxygen saturation, fractional oxyhemoglobin, carboxyhemoglobin and methemoglobin. Crit Care International, April–May, 8-9, 1995.
Moran RF, The case for standardized terminology: Oxygen "saturation" values can trick the unwary and lead to clinical misjudgement., Crit Care Med, 21:5, 805-807, 1993.
Moran RF, Lab Consultant: [High Percentage of zero carboxyhemoglobins due to correction algorithm for small, "impossible", values.] Clin Chem News, 18:12, 18-19, 1992.
CLSI document C25A, can provide in-depth information and references.
- ↑ Rodkey FL, Hill TA, Pitts LL, Robertson RF (August 1979). "Spectrophotometric measurement of carboxyhemoglobin and methemoglobin in blood". Clinical Chemistry. 25 (8): 1388–93. PMID 455674. Retrieved 2009-07-17.
- ↑ Rees PJ, Chilvers C, Clark TJ (January 1980). "Evaluation of methods used to estimate inhaled dose of carbon monoxide". Thorax. 35 (1): 47–51. doi:10.1136/thx.35.1.47. PMC 471219. PMID 7361284. Retrieved 2009-07-17.
- ↑ Coulange M, Barthelemy A, Hug F, Thierry AL, De Haro L (2008). "Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning". Undersea & Hyperbaric Medicine. 35 (2): 107–11. PMID 18500075. Retrieved 2009-07-17.
- ↑ Vegfors M, Lennmarken C (May 1991). "Carboxyhaemoglobinaemia and pulse oximetry". British Journal of Anaesthesia. 66 (5): 625–6. doi:10.1093/bja/66.5.625. PMID 2031826.